The quick-reference dosimetry sheet our students keep next to the console: beam parameters, isodose concepts, brachytherapy isotope data, and the calculation formulas that come up on every plan. Here's a preview — the full printable PDF covers it all, plus OAR constraints, prescription tables, and organ tolerance limits.

Free printable PDF

Get the full Dosimetry Cheat Sheet (PDF)

Enter your email and the complete printable cheat sheet downloads instantly — every formula, isotope, and tolerance table in one reference.

Beam parameters

Typical megavoltage buildup depths and depth-dose values for a 10×10 cm field.

Energydmax depthPDD at 10 cm
6 MV~1.5 cm~67%
10 MV~2.5 cm~74%
15 MV~3.0 cm~77%
18 MV~3.5 cm~80%

Isodose concepts

  • Isodose curves connect points of equal dose; dose builds up to dmax, then falls with depth.
  • Beam edges bow outward from divergence and scatter, so penumbra widens with depth.
  • A wedge tilts the isodose lines toward its thin end — the thick end absorbs more dose.
  • On plan review, the 95% isodose line should cover the PTV and the 105% line should stay inside the target.
  • Multi-beam plans stack isodose distributions; hot spots form where beams intersect.

Brachytherapy isotope table

Average energies and half-lives for the sources you see in clinic and on exams.

SourceAvg energyHalf-life
Co-601.25 MeV5.27 y
Cs-1370.662 MeV30.0 y
Ir-192~0.38 MeV73.8 d
I-12528 keV59.4 d
Pd-10321 keV17.0 d
Ra-2235.0–7.5 MeV (α)11.4 d

Calculation formulas

A sample of the formulas inside the full sheet.

Inverse square law

I₂ = I₁ · (d₁/d₂)²

Dose rate vs distance from a point source.

Equivalent square

eq = 4 · A / P

Field size for dosimetry-table lookups.

Monitor units (SAD)

MU = D / (K · Sc · Sp · TMR · TF)

MU for an isocentric (SAD) setup.

Monitor units (SSD)

MU = D / (K · Sc · Sp · PDD · TF · f_ISL)

MU for a fixed SSD setup; f_ISL = (SSD/SSD+d)².

Inverse-square correction

f_ISL = (SAD₁ / SAD₂)²

Correct dose-per-MU when source distance changes.

Wedge transmission

TW = D_wedge / D_open

Wedge factor / transmission to apply.

Linear-quadratic survival

S = exp[−(α·D + β·D²)]

Cell survival vs single fraction dose D.

Biologically effective dose

BED = n·d · (1 + d / (α/β))

Biological effect for n fractions of size d.

EQD2

EQD2 = D · (d + α/β) / (2 + α/β)

Convert a regimen to 2-Gy equivalent dose.

Half-life

T½ = ln(2) / λ

Relates decay constant to half-life.