CT is the backbone of RT — it provides the 3D anatomy and the density map used for treatment planning and dose calculation.
Imaging in the RT workflow
Medical imaging
The imaging modalities, accessory equipment, body planes, and digital-imaging concepts that drive radiation therapy simulation, planning, and verification. Tap a card to expand the detail.
Exposure parameters: mA, kV & mAs
How changing each factor affects the CT image, the DRR, and patient dose
kV (tube voltage)
- Effect on CT image
- Higher kV → more penetrating (higher-energy) beam, more Compton / less photoelectric, so image contrast drops slightly and beam-hardening artifacts lessen; HU stays calibrated. Lower kV raises contrast (bone vs soft tissue) but increases noise and dose for the same mAs.
- Effect on DRR
- DRR contrast follows the CT: a lower-kV planning CT gives stronger bone–soft-tissue contrast in the DRR for portal matching, while a higher-kV CT is smoother and more penetrating but flatter.
- Effect on patient dose
- Higher kV lowers surface dose and improves penetration for the same mAs; raising kV without adjusting mAs usually reduces patient dose.
mA (tube current)
- Effect on CT image
- Higher mA delivers more photons per rotation, reducing quantum mottle/noise and improving SNR and low-contrast detectability; spatial resolution and contrast are otherwise unchanged.
- Effect on DRR
- A higher-mA CT produces a less noisy DRR with crisper bony edges, improving the accuracy of DRR-to-portal/CBCT registration.
- Effect on patient dose
- Patient dose scales linearly with mA — doubling mA doubles dose while reducing noise by ~√2.
mAs (mA × exposure time)
- Effect on CT image
- mAs sets the total photons per slice/projection. Image noise ∝ 1/√mAs, so increasing mAs reduces noise and improves image quality; lowering mAs increases noise and streak artifacts.
- Effect on DRR
- A low-mAs (or dose-sparing) CT yields a grainy DRR that is harder to match; a higher-mAs CT yields a sharper, cleaner DRR reference.
- Effect on patient dose
- Dose is directly proportional to mAs — the main lever for CT dose (CTDIvol/DLP) without changing contrast.
A DRR is a synthetic 2D image computed from the planning CT, so its quality and contrast are inherited from the CT acquisition settings — it is not re-acquired at the linac.
14 topics
A cone-beam CT mounted on the linac gantry produces a 3D volume of the patient in treatment position for daily setup verification.
MRI offers superior soft-tissue contrast without ionizing radiation — key for brain, head & neck, prostate, and gynecologic targets.
PET images metabolism (commonly FDG uptake) to support staging, targeting, and response assessment.
Other modalities fill niches from real-time localization to motion-resolved planning and planar verification.
Contrast improves target visualization; fiducials provide a fixed, imageable reference inside the patient for setup and gating.
Devices hold the patient reproducibly and manage internal motion so the image and the treatment match.
Detectors and anti-scatter devices determine image quality; surface imaging adds real-time, marker-less monitoring.
Three orthogonal planes define how cross-sectional images are described and reformatted in RT.
Standard terms and the DICOM coordinate system describe location and orientation unambiguously.
DICOM is the medical-imaging data standard; DICOM-RT extends it for the objects that flow between scanner, TPS, and record system.
Storage systems and review tools manage the digital image lifecycle from acquisition to contouring.
Registration aligns datasets; segmentation defines the volumes used for planning.
The recurring terms that connect imaging to planning, delivery, and verification.